MD-120 Auto Chemistry Analyzer

Image

System Function

  • Category: Random access, fully automatic chemistry analyzer
  • Throughput: Constant throughput 100 tests/hour
  • Methodology: End point, Fixed Time, Kinetic
  • Reagent: Closed or Open system
  • Reagent Position: 20–40 positions (20 fixed + 20 interchangeable)
  • Sample Position: 20 positions (interchangeable)
  • Minimum Reaction Volume: 150µL
  • Carryover: < 0.01%
  • Water Consumption: < 3.5 L/Hour
  • Maximum Reaction Time: 15.6 minutes (Single reagent), 10.2 minutes (Dual reagent)
  • Maximum Reaction Volume: 400µL
  • Principle: Colorimetry, Turbidimetry

Technical Excellence

• Global Top-Tier Components

Guarantees accurate performance and long-term durability

• Efficient Silent Water Cooling

Ensures lamp stability and reagent cooling efficiency, while minimizing dust ingress into the instrument.

• Sealed Optical Path

Exceptional optical steadiness, maintenance-free operation

• Verified Performance

Rigorously validated reliability across 100+ reagents


Key Features

• Compact & Efficient

Space-saving design with throughput of 100 tests/hour

• One-Step Batch Scanning

Effortlessly track all reagents and samples with 100% code capture

• Intelligent Refrigeration

24-hour continuous cooling with 40-position tray (20 fixed + 20 customizable)

• Auto-Cleaning

4-stage, 8-step cuvette cleaning for 50 reusable UV-transmitting cuvettes | Automated cleaning solution delivery

• Precision Sampling

Liquid level sensing | 3D collision avoidance | Automatic bubble discharge

• High-Speed Vortex Mixing

Achieve Zero Sedimentation, Zero Stratification

• Smart Operation

10.1" industrial touchscreen | Intuitive and Advanced Capabilities Software


Sample Reagent Probe & Mixer

  • Sample Mode: Random access, STAT sample priority
  • Sample Volume: 2–50µL, step by 0.1µL
  • Sample Disk: 20 positions (interchangeable)
  • Sample/Reagent Probe: One probe for sample and reagent with liquid level detection, collision protection and reagent volume real time monitoring function
  • Sample Tube: Regular sample tube, Vacuum tube, Eppendorf tube, plastic tube, etc.

Reaction System

  • Reagent Volume: R1 150–450µL, R2 10–300µL, step by 1µL
  • Reagent Disk: 20–40 positions (20 fixed + 20 interchangeable), 24 hours cooling, 4–12°C
  • Reagent Bottles: Compatible with Olympus and Hitachi reagent bottles
  • Mixer: High polished nano material mixer (Tentative)
  • Reaction Cuvettes: 50 cuvettes, semi-permanent rigid UV special plastic cuvettes
  • Wash Station: 4-cleaning needles auto wash station
  • Reaction Temperature: 37±0.1°C

Optical System

  • Heater: Metal thermostat for reaction disk
  • Lamp: Halogen Tungsten lamp, about 2000 hours service life
  • Filter: FMSS (Full-sealed Matrix Spectrometric System)
  • Wavelength: 8 wavelengths 340, 405, 450, 510, 546, 578, 650, 670nm (800nm, 700nm optional)
  • Resolution: 0.0001Abs
  • Linearity Range: 0.5–5Abs (540nm), 0–4.0Abs (510nm)
  • Accuracy: 0.5Abs < ±0.02Abs, 1.0Abs = 0.04Abs
  • Stray Light: ≥5
  • Stability: <0.01Abs/10 minutes
  • CV: <1.5%
  • Half Band Accuracy: ≤2nm

Cal & QC

  • Detector: Photodiode detector array
  • Calibration Mode: One point linear, two point linear, multi-point linear, Logit-4P, Logit-5P, spline, exponential, polynomial
  • Calibration Curve: Calibration curve auto check, auto curve fitting
  • QC Rule: Westgard multi-rule, Cumulative sum check, Twin Plot
  • QC Warning: Out of control auto warning, data automatic record and analysis

Software

  • Operating System: Windows 11
  • Data Storage: Decided by computer hard disc capacity
  • LIS Interface: Bi-LIS interface
  • Print Mode: Multi default formats, user-defined formats
  • System Monitor: Real time monitoring for sample disk, reagent disk, reaction disk
  • Real time monitoring for QC status.
  • Real time monitoring for reaction cuvettes status, lamp status and temperature.
  • Real time monitoring for reagent volume, reaction curve, calibration curve and QC curve.
  • Linearity range limitation, substrate exhaustion judgement and prozone detecting.
  • Abnormal status warning.

Other Function

User authority setting, test panel function, calculated/manual parameters programming, carryover setting, sample and reagent blank auto calculation, automatic failure recovery, automatic print, data statistic, auto/manual dilution test, auto retest.


Others

  • PC Configuration: Embedded computer
  • Monitor: 10.1-Inch touch screen
  • Dimension: 595(L) × 548(W) × 583(H)mm
  • Power Supply: AC 100–240V, 50/60Hz ±1Hz, ≤350VA
  • Net Weight: 40Kg
  • Optional Parts: Barcode reader, External printer

Google Map

PT. Radhar Delta Bersaudara

Medical, Hospital & Laboratory Equipment

Komplek Taman Candiloka Blok E2 No. 43 Candi, Sidoarjo, Jawa Timur 61271

Kontak Kami

+62 31-99717-7767

+62 81-13344-957

ptradhar.db@gmail.com

© 2025 PT. Radhar Delta Bersaudara. All Rights Reserved.

Developed by Kahfidea.id. Template by HTML Codex